www.pdr-rework.com
Maintenance, Troubleshooting and Calibration
There are 4 sections in this area of the manual
Main Equipment Maintenance
Calibration Procedures
General Troubleshooting
Warranty
Main Equipment Maintenance
All maintenance should be performed when the machine is switched off, and has been allowed to cool to avoid the
operator accidentally receiving burns from hot parts of the machine.
Lubrication
The only lubrication necessary is the application of a thin coating of grease to the threads of the height adjuster and
the lens attachment. The recommended grease is a Lithium grease. After application all surplus grease should be
removed by wiping with a clean tissue. Bearings should not need lubrication.
Cleaning
At the end of each working day wipe any surplus flux and/or solder paste from the equipment with a clean
cloth.
Periodically clean the exposed surfaces of the equipment with a soft cloth slightly moistened with water.
The lens assembly should be cleaned using Iso-Propylalcohol (IPA). Care should be taken to ensure that the lens
cloth is clean as any contamination of the cloth could cause scratching of the glass surfaces.
Caution: do not use solvents for cleaning any part of the equipment as they may damage the operation and finish.
Changing Fuses
Fuses should only be changed with the equipment switched off, and the equipment isolated from the electrical
supply by removing the IEC socket from the machine.
The mains fuse is located at the back of the equipment and is changed by withdrawing the fuse holder and
withdrawing the fuse.
Changing the IR Lamp
The IR lamp may be changed by the following procedure:
Ensure that the equipment is switched off and mains cable is removed from the mains supply.
Slacken the small screw securing the lamp housing to the lens unit. Lift off housing.