6.3.8 Measuring AC Current
Caution!
Do not make current measurements in circuits with a potential over 400 V. Extreme risk of injury and / or the risk of damage
to the meter.
Warning!
* Do not connect voltages directly across the terminals. The device must only be connected in series with the circuit
to be measured.
* The 10 A input is protected by an fuse corresponding. When connecting a voltage source to this input may result in
injury and danger of damage to the unit.
To measure an AC current which is less than 400 mA, do the following:
1.) Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is mA. mA and 10A
will display on the right bottom of screen, press F4 or F5 to switch the measurement between mA and 10A. 400mA
is acquiescently.
2.) Press the SET key once and ACA is visible at the top of the screen.
3.) Insert the black lead into the COM banana jack input and the red lead into the mA banana jack input.
4.) Connect the red and black leads to the measured point and the AC current value of the measured point will be
displayed on the screen.
Look at the display you can see a screen that looks like the following figure 25.
Figure 25: AC Current Measurement for 400 mA
-89-