To measure an AC current which is larger than 400 mA, do the following:
1.) Press the A key, an instruction prompt you to plug the multimeter test leads in the right position. After check, press
any key to enter into multimeter measure and DCA appears at the top of the screen. The unit on the main reading
screen is mA.
2.) Press the SET key once and ACA is visible on the top of the screen.
3.) Press F5 to select 10A measure, the unit of main reading window is A.
4.) Insert the black lead into the COM banana jack input and the red lead into the 10A banana jack input.
5.) Connect the red and black leads to the measured point and the AC current value of the measured point will be
displayed on the screen.
6.) Press F4 return to 400mA measure.
Look at the display you can see a screen that looks like the following figure 26.
Figure 26: AC Current Measurement for 10A
-90-