3. Specifications
Display
LCD with automatic backlight
Measurement
Range
2000 Lux
2000 to 20000 Lux
20000 to 200000 Lux
Resolution:
1 Lux (2000)
10 Lux (20000)
100 Lux (200000)
Sample rate
1.5 x Second, typically
Accuracy
< 10k Lux
+/- 4 % rdg.
> 10k Lux
+/- 5 % full Scale
calibrated to standard incandescent
lamp at color temperature 2.856 K
Low Battery
Indication
Symbol “ “ shown in the display
Auto Power Off
After 10 Minutes
DATA-HOLD
Holds measurement value in display
Power Supply
9V battery, 6F22 or equivalent
Operating
Environment
0°C ~ 50°C; <80% RH
Storage
Environment
-10°C ~ 50°C; <70% RH
Size (WxHxD)
55 x 155 x 35mm
Weight
120g