CALIBRATION: “SEL.CAL” Use Gross/Net to enter the menu indicated by a flashing “C” on
the left and live weight is displayed. Scale zero (dead load) or adjusting span (single or multi-
point) are independent. Therefore either can be done and repeated as necessary before
exciting calibration. If an error has been made, exit without “storing” will return to prior setup.
KEY (FUNCTION) DISPLAY Definition
(Live weight 123 lb) “C”__123 Cal mode scale reading
Zero (acquire dead load) “------“ to “C___0.0” acquires new dead load
(live weight 5000 lb) “C”__4995 Scale reading with load
Enter numeric value directly:
(Adjustment complete) 005000 adjusted value
Then ENT:
(adjust span) “------“ to “C” 5000 displays new span
Repeat as required then ENT to exit CAL
“Save ?” “No” or “Save ?” “Yes” use Units to select and ENT to store “yes” with changes or
“no” to exit without changes.
Continue with ENT to “Ent AC” which allows access code change by entering a new four digit
code and ENT or ENT with no entry to maintain current password.
Option 5 Ten point calibration: Allows up to 10 span points (pt1…….pt10). Zeroing the scale
clears the existing values. Points are assigned incrementally with error indication if the addition
is not above the prior point or exceeding scale capacity.
Filter selection included for rolling or box averaging.
5.1 OFF, On Enable 10 point span
5.2 A, b
A : Rolling average
B : Box average