D04-IFU-0001400 11 Revision: B
This device is compatible with STERRAD
®
100S when following the
sterilant manufacturer’s procedures and guidelines.
3-3 Use
Caution
Thoroughly inspect the device before each use.
Check the
function of the device and
visually inspect optics for moisture or
contamination. Do not drop
the device or subject it to sudden
impact.
Refer to section 5,
Warranty
for information about
damage to the device.
Caution
Never insert the
rigid Laryngostroboscope in water or any other
substance or agent with a temperature greater than 150 ºF (65
ºC). Hot beads can damage the
rigid Laryngostroboscope and
should not be used for anti
-fogging.
Note
The optimal working distance from the distal window for
vi
sual use of this rigid Laryngostroboscope is 5 to 50 mm.
Note
The angle of entry with the 70-degree rigid
Laryngostroboscope
is markedly different from the 90-degree
rigid
Laryngostroboscope to which you may be accustomed.
Instead of the
rigid Laryngostroboscope pointing upward in the
patient’s mouth, the 70
-degree rigid Laryngostroboscope angles
downward to allow a closer view
of the vocal fold.
Warning
Although complications resulting from use of a rigid
Laryngostroboscope
for viewing the larynx are extremely rare,
the following risks are possible:
Laryngospasm.
Blood pressure elevation during procedure.
Bleeding as a result of injury caused by improper scoping
procedure.
Adverse reaction to anesthetic (if used to prevent gagging).