D04-IFU-0001400 11 Revision: B
• This device is compatible with STERRAD
®
100S when following the
sterilant manufacturer’s procedures and guidelines.
3-3 Use
Caution
Thoroughly inspect the device before each use.
Check the
function of the device and
visually inspect optics for moisture or
contamination. Do not drop
the device or subject it to sudden
Refer to section 5,
Warranty
for information about
rigid Laryngostroboscope in water or any other
substance or agent with a temperature greater than 150 ºF (65
ºC). Hot beads can damage the
rigid Laryngostroboscope and
should not be used for anti
-fogging.
The optimal working distance from the distal window for
sual use of this rigid Laryngostroboscope is 5 to 50 mm.
The angle of entry with the 70-degree rigid
is markedly different from the 90-degree
Laryngostroboscope to which you may be accustomed.
rigid Laryngostroboscope pointing upward in the
-degree rigid Laryngostroboscope angles
downward to allow a closer view
of the vocal fold.
Warning
Although complications resulting from use of a rigid
Laryngostroboscope
for viewing the larynx are extremely rare,
the following risks are possible:
Laryngospasm.
Blood pressure elevation during procedure.
Bleeding as a result of injury caused by improper scoping
procedure.
Adverse reaction to anesthetic (if used to prevent gagging).