10 PERCUSSIONAIRE
®
Disinfection solution:
Use standard household bleach (sodium hypochlorite 5.25%) for disinfection.
Mix solution of 1-part household bleach and 8-parts water. Example: 8 oz
bleach to 64 oz water or 250 ml bleach to 2 L water. (Review and follow all
safety precautions printed on bleach container).
Disinfection procedure:
1. Wear disposable gloves.
2. Pour bleach solution into an airtight container large enough to hold 2
liters or 64 ounces.
3. Completely immerse all the Phasitron® parts and tubing harness
(including mouthpiece, mask, and other accessories) into bleach solution.
Do not disinfect antibacterial filter.
4. Let soak for 10 minutes minimum. Do not exceed one hour.
5. Rinse o cold-disinfection solution using sterile water or filtered (less
than or equal to 0.2-micron filter) water. Do not use tap water.
6. Gently shake all parts to remove as much water as possible, then dry with
clean, lint-free cloth or paper towel.
7. Place on clean, lint-free cloth or paper towels and allow to air dry
completely.
8. Insert all parts, not assembled, into sealable plastic bag and store in a dry,
clean location.