Institutional and Homecare
Cleaning Manual
Percussionaire® Phasitron® & Nebulizer
Question and Answer IconNeed help?

Do you have a question about the Percussionaire Phasitron and is the answer not in the manual?

Summary

Percussionaire Controller Device Cleaning

Controller Device Cleaning Frequency

Cold cleaning and disinfection of the controller & display device surface should be performed weekly, or as needed.

Phasitron Kit and Harness Cleaning

Phasitron Kit Cleaning Schedule

Basic cleaning after each treatment. Cold disinfection to be performed weekly after regular basic cleaning.

Tubing Harness Cleaning

Tubing Harness Cleaning Schedule

Cold cleaning and disinfection of the tubing harness to be performed weekly.

Safety Precautions and Important Safeguards

Cleaning Procedures: After Each Use

Phasitron Disassembly

Steps for disassembling the Phasitron unit, including separating the nebulizer and removing the bowl from the cap.

Phasitron Parts and Accessories Cleaning

Rinsing and Washing Steps

Rinse parts, wash with fragrance-free soap, and use a soft brush for stubborn soil. Do not wash antibacterial filter.

Drying Parts and Tubing

Shake parts to remove water, dry with lint-free cloth, and air dry. Wipe tubing harness with soapy water and rinse.

Disinfection Solution and Procedure

Disinfection Solution Preparation

Mix 1 part household bleach (5.25%) with 8 parts water. Review safety precautions.

Disinfection Steps

Immerse parts in bleach solution for 10 mins to 1 hour. Rinse with sterile/filtered water, then dry and store.

Weekly Lubrication

Lubrication Process

Coat O-rings and specific components with approved lubricant before re-assembly and lubrication.

Reassembly of Phasitron and Tubing Harness

Nebulizer Bowl and Venturi Assembly

Twist nebulizer bowl into cap until it locks. Ensure inner stem on diaphragm is pressed firmly into the venturi.

Phasitron Reassembly Steps

Venturi O-ring Inspection and Insertion

Inspect O-ring on sliding Venturi for damage. Insert Venturi and spring into Phasitron body.

Phasitron Reassembly Connections

White Cap and Patient Interface Attachment

Screw white cap onto Phasitron body clockwise. Attach mouthpiece, mask, or corrugated tubing.

Drying Inside the Tubing

Tubing Drying Method

Use air from the controller device to dry tubing interiors. Ensure check valve arrow points away from the controller.

Reattaching Tubing Harness to Phasitron

Connecting Tubing to Nebulizer and Filter

Connect yellow tubing to nebulizer bowl. Ensure check valve direction. Connect red tubing to bacterial filter.

Phasitron Tubing Reattachment

Connecting Tubing to Phasitron Body and Nebulizer

Connect red tubing to Phasitron front, green tubing to nebulizer top, and white tubing to Phasitron rear.

Reattaching Tubing Harness to Controller Device

Controller Device Tubing Connections

Connect red, white, green, and yellow tubing to their corresponding ports on the Percussionaire controller device.

Surface Disinfection of Controller Device

Controller Device Cleaning Method

Use an approved wipe or spray agent onto a damp cloth to wipe the device surface. Do not submerge.

Percussionaire Phasitron Specifications

General IconGeneral
PortabilityPortable
Patient InterfaceMask, Mouthpiece, Tracheostomy Adapter
IndicationsSecretion Mobilization, Atelectasis
ContraindicationsUntreated pneumothorax