The replaced tube should be able to withstand at least 10℃ higher than the experimental
temperature.
2. Fix with Teflon tape.
Firmly fix the tube by winding Teflon tape around the connection.
When the Multi-Cell does not move
1. Check with the connector port.
Check if the Multicell Holder connector is connected firmly to the Lambda 365.
When the intensity value is low
1. Recalibrate the Multi-Cell.
Recalibrate if the light beam does not go through the center of the cell holder.
2. Replace the lamp.
If the intensity value is still low after the recalibration, the lamps of the light source for
measurement may be deteriorated and need replacing. Contact PerkinElmer for replacing the
lamp.
When the Multi-Cell does not reset
If resetting fail, the photo interrupter switch inside the Multi-Cell will need replacing. Contact your
PerkinElmer Service representative.