8. Select Save Result to save the result. If the following message box appears, select OK.
Troubleshooting
When liquid leakage occurs
1. When liquid is leaking at the fittings.
Using wrench, turn the fitting counter-clockwise to detach from the Water Jacketed Multi-Cell
Holder.
Remove the Teflon which is wound around the fitting. Wind a length of new Teflon tape
around the fitting.
Turn the fitting clockwise to attach it to the Water Jacketed Multi-Cell Holder.
2. When liquid is leaking between the quick coupling mount (or fitting) and the tube.
Replace the tube.
The replaced tube should be able to withstand at least 10 ℃ higher than the experimental
temperature.
3. When liquid is leaking between the quick coupling mount and the Water Jacketed Multi-Cell
Holder.
Contact PerkinElmer since the quick coupling mount needs to be replaced.
When constant temperature is not maintained
1. Check if the circulator is working normally.
Check if the temperature is maintained normally in the circulator.
Check if liquid is transferred normally to the Water Jacketed Multi-Cell Holder through the tube.
2. Check the tube connection.
Check if the tube is connected tightly.
When the tube frequently becomes detached
1. Replace the tube.
You should replace the tube as it is aged.