Wavelength Range | 190 - 1100 nm |
---|---|
Power Requirements | 100-240 VAC, 50/60 Hz |
Photometric Accuracy | ±0.002 Abs (at 1 Abs) |
Baseline Stability | < 0.0005 Abs/hour |
Light Source | Deuterium and Tungsten-Halogen lamps |
Display | Touchscreen |
Photometric Repeatability | ± 0.001 Abs |
Information on ordering supplies, replacement parts, and accessories from PerkinElmer, including website and contact numbers.
Highlights key attributes of the multi-cell holder, such as size, durability, and temperature control.
Provides physical measurements for the outline and inner dimensions of the multi-cell holder, along with its weight.
Details technical specifications including space between cells, moving distances, and moving times.
Specifies the operating temperature range and method for temperature control using a circulator.
Guidance on diagnosing and resolving issues related to liquid leaks at fittings, between mounts and tubes, or holders.
Steps to check circulator function, temperature maintenance, and tube connections for consistent temperature.
Recommendations for fixing tube detachment issues, including replacing aged tubes and using Teflon tape.
Troubleshooting steps for when the multi-cell holder fails to move, focusing on connector port checks.
Instructions for recalibrating the multi-cell or replacing the lamp if low intensity values persist.
Guidance for resetting failures, suggesting replacement of the photo interrupter switch.