09931251D
7
10. Turn on the power switch of the instruments.
Measurement
NOTE:
Before executing System Self Test,the Water Jacketed 8-Position Multi-Cell Holder has to be installed first.
Otherwise, the instrument can be damaged electrically and do not operate properly.
1. Install the Water Jacketed 8-Position Multi-Cell Holder.
2. Close the sample compartment cover and turn on the instrument.
CAUTION
Do
not
start the UV Express software before finishing the initialization of grating.
NOTE:
Start the System Self Test after warming up the system for at least 20 minutes.
3. Double-click on the UV Express Software folder and select one of the mode for starting.
4. The following message box will be shown. Empty the cell holder and close the lid firmly. Click OK.
5. System Self Test stars. Click OK after finishing the System Self Test.
6. Select New to open a new window. Select Experiment Type and select OK.
7. Open Method in the main menu.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Perkin-Elmer Lambda 365 and is the answer not in the manual?

Perkin-Elmer Lambda 365 Specifications

General IconGeneral
Wavelength Range190 - 1100 nm
Power Requirements100-240 VAC, 50/60 Hz
Photometric Accuracy±0.002 Abs (at 1 Abs)
Baseline Stability< 0.0005 Abs/hour
Light SourceDeuterium and Tungsten-Halogen lamps
DisplayTouchscreen
Photometric Repeatability± 0.001 Abs