EasyManua.ls Logo

PERSEE T6 U - Page 24

PERSEE T6 U
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
In the case of system set, press 【▲】 and 【▼】 to move the cursor to the option of Reset System,
then press ENTER to gain access to the menu of reset system.
(Fig. 3-20)
In the case of reset system, press 【▲】 and 【▼】 to move the cursor to the required option, then
press ENTER.
The original parameters setting:
Module Style: 8-cell changer
Cell Number: 5
Reagent Blank: No
Cell Blank: No
Change Lamp Wavelength: 359.9nm
Dark Current: reset
The instrument will remember the parameters, which is set according to the users requirements. It is
not recommended to reset system.
Input Change Lamp Wavelength
In the case of system set, press 【▲】 and 【▼】 to move the cursor to the Change Lamp WL, then
press ENTER to input value.
(Fig. 3-21)
At the bottom of this page, enter the wavelength by pressing the numeric keys (0~9) and decimal point
【.. Press RETURN key to clear the mistakes, and press CE to clear step by step, then press
ENTER key to confirm . Press ENTER key to confirm and quit. If nothing is to be entered, press
RETURN key to quit this page directly. A 3-digit figure entry is allowed. The range is from 300 to 400.
If the entered figure exceeds the range and you have pressed ENTER, the system will delete what has
been entered automatically and you should start a new entry again.