21
Chapter 4 Specifications
Wavelength
Range
190nm-1100nm
Wavelength
Accuracy
+1.0nm (with automatic wavelength correction)
Wavelength
Reproducibility
0.2nm
Spectral
Bandwidth
2 nm
Photometric
Accuracy
+0.3 %T
Photometric
Reproducibility
0.15 %T
Stray Light
0.05 %T220nm NaI,340nm NaNo2
Baseline
Flatness
0.002A1000-200nmafter 1 hour warm-up
Noise
0%T+0.05%T (500nm)
100%T:≤0.15%T (500nm)
Baseline
Stability
<0.001A/1h (500nmafter 1 hour warm-up)
Photometric
Range
Abs-0.3-3Abs
Transmittance0-200%T
Interface
PC interface via RS232 link
Printer interface via parallel output port
Optional
Program Card
With the extended interface for optional
program cards
Light Source
12V20W tungsten lampdeuterium lampwith
flange
LCD
128X64 backlight digital LCD
Sample
Compartment
Inner Dimensions235X130X110mm
The maximum pathlength: 100mm
Build-in automatic multiple cells
Detector
Silicon photodiode
Photometric
System
The split-beam monitoring ratio system
Voltage
230V/120V (original setup)
Weight
11kg
Dimensions
476X362 X 225mm