EasyManua.ls Logo

PERSEE T6 U - Page 30

PERSEE T6 U
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
Set the change lamp wavelength, the instrument will switch over the lamp automatically.
Why isnt the stray light eligible?
The poor power and the resultant raise of the dark current adversely affect it. Perform dark current
correction will solve the problem. In addition, if the instrument is placed in the humid environment for a
long time, it will also cause the stray light raises higher.
It is recommended that switch on the power of the instrument once a week to avoid some problems
caused by leaving unused.
How to clear the data?
T6U is designed to clear all the data after printout. The system will promote the user to print data after
changing the parameters. If the user selects Dont print, the system will also clear all the data.