EasyManuals Logo

pg instruments T60U User Manual

Default Icon
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
19
In the case of system set, press ǏƷǐ and Ǐǐ to move the cursor to the option of Reset System,
then press ǏENTERǐ to gain access to the menu of reset system.
(Fig. 3-20)
In the case of reset system, press ǏƷǐ and Ǐǐ to move the cursor to the required option, then
press ǏENTERǐ.
The original parameters setting:
Module Style: 8-cell changer
Cell Number: 5
Reagent Blank: No
Cell Blank: No
Change Lamp Wavelength: 359.9nm
Dark Current: reset
The instrument will remember the parameters, which is set according to the users requirements. It is
not recommended to reset system.
Ĵ
Input Change Lamp Wavelength
In the case of system set, press ǏƷǐ and Ǐǐ to move the cursor to the Change Lamp WL, then
press ǏENTERǐ to input value.
(Fig. 3-21)
At the bottom of this page, enter the wavelength by pressing the numeric keys (0~9) and decimal point
Ǐˊǐ. Press ǏRETURNǐ key to clear the mistakes, and press ǏCEǐ to clear step by step, then press
ǏENTERǐ key to confirm . Press ǏENTERǐ key to confirm and quit. If nothing is to be entered, press
ǏRETURNǐ key to quit this page directly. A 3-digit figure entry is allowed. The range is from 300 to 400.
If the entered figure exceeds the range and you have pressed ǏENTERǐ, the system will delete what has
been entered automatically and you should start a new entry again.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the pg instruments T60U and is the answer not in the manual?

pg instruments T60U Specifications

General IconGeneral
Brandpg instruments
ModelT60U
CategoryMeasuring Instruments
LanguageEnglish