6CP100/50-10
5-2
OPERATIONAL VERIFICATION PROCEDURE
14. Move short from 50 kV output to 100 kV output and ground. Repeat step 13. Unit should trip off at
approximately 10 mA.
NOTE: Do not attempt to check 110% level of current trip under shorted output conditions. Rated output
currents are only available through a capacitive load of sufficient value. Shorted current values should
never exceed 50% of rated currents for the High Voltage Tap being used. If desired to test at 110% of
current rating, the 50kV/200mA tap must be connected to a High Voltage Capacitor of approximately 6.5
nanofarads rated at 50 kVAC or higher (for 50 Hz).
15. Turn Off Main Power Circuit Breaker on Control Panel.
END OF OPERATIONAL VERIFICATION TEST!