EasyManua.ls Logo

Phenix 4100-10 - Calibration Procedures; Output Voltmeter Calibration

Phenix 4100-10
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-1
SECTION 6: CALIBRATION
All calibrations have been done at the factory. Periodic calibration of the output voltmeter and output
currentmeter should be done annually.
NOTE: Refer to Electrical Diagram Section for schematics pertaining to the model number of your test set.
Locating the Calibration Adjustments
The calibration points are shown in the following diagram.
1. Output Voltmeter
Connect a precision high voltage voltmeter across the output to ground on low range. Raise the output to
approximately 80% of range. Adjust the reading on the panel meter (M1) by means of potentiometer R1
to a corresponding reading. Check Linearity and calibration at various other main points of the range
such as 20%, 40%, 60%, 100%. Repeat for High range, adjusting R2.
R101
10%
Ovld
R103
110%
Ovld
R235
R201
20mA
CM
R202
2mA
CM
R203
200uA
CM
R204
20uA
CM
R2
100kV
VM
R1
20kV
VM
CAUTION:
Calibration should only be done by persons familiar with High Voltage testing and safety
procedures.

Related product manuals