Do you have a question about the PHILIPS Respironics DreamStation BiPAP S/T and is the answer not in the manual?
Device Type | BiPAP S/T |
---|---|
Manufacturer | PHILIPS Respironics |
Model | DreamStation BiPAP S/T |
Humidification | Integrated heated humidifier |
Data Storage | SD card |
Altitude Compensation | Automatic |
Heated Tubing | Optional |
Warranty | 2 years |
Compliance Monitoring | Yes |
Modes | S/T, CPAP |
Weight | 3.5 lbs |
Power Supply | 100-240V, 50/60Hz |
Ramp Time | 0 to 45 minutes |
Rise Time | 1-5 |
Operating Conditions | 5° to 35° C (41° to 95° F) |
Alarms | Power failure |
Pressure Range | 4-25 cm H2O |
Accessing and adjusting advanced therapy modes, pressure, AVAPS, and AAM settings.
Procedures to verify the proper functioning of the device's pressure and alarm systems.
Steps to verify the pressure output of the BiPAP S/T or AVAPS device using a manometer.
Procedures for testing various device alarms like Patient Disconnect and Apnea alarms.
Testing the patient disconnect alarm for proper activation and timing.
Testing the apnea alarm to ensure it triggers correctly based on patient breathing.
Lists and describes various device error levels, codes, and troubleshooting actions.
Guides troubleshooting based on common problems like no power or airflow issues.
Detailed steps for removing and installing the Printed Circuit Assembly (PCA).
Procedures for replacing the blower, its housing, and the rear panel.
Provides procedures for repairing and replacing humidifier components.
Lists all necessary hardware and software for performing device testing.
Step-by-step guide for executing the final testing procedure using the Test Executive.