12 Specifications
227
1
The power loss indication functionality of the Nurse Call Relay board is not supported with fetal
monitors.
Performance Specifications
NOTE
Your monitor's default settings can be permanently changed in Configuration Mode.
The default settings specified here refer to the settings initially shipped with the monitor.
Fetal Monitors
RS232 (Standard) Connectors RJ45 (8-pin)
Power none
Insulation Basic insulation (reference voltage: 250 V; test voltage: 1500 V)
RS232 (Independent display
interface option)
Connectors RJ45 (8-pin)
Power none
Isolation none
Flexible Nurse Call Relay
1
Connectors 20 pin MDR (Mini D-Ribbon), active open and closed contacts
Contact <=100 mA, <=24 V DC
Isolation Basic insulation (reference voltage: 250 V; test voltage: 1500 V)
Delay <[Configured Latency +0.5] sec
Interface Specifications
Fetal Monitor Performance Specifications
Alarm Signal System alarm delay.
The system alarm delay is the processing
time the system needs for any alarm to be
indicated on the fetal monitor, after the
measurement has triggered the alarm.
less than 4 seconds
Pause duration 1, 2, 3 minutes or infinite, depending on the
configuration
Extended alarm pause 5 to 10 minutes
Sound pressure range min. 0 dB(A) max. 45—85 dB(A)
Review Alarms Information all alarms/INOPs, main alarms on/off, alarm
silence and time of occurrence
Capacity 300 items
Real time Clock Range from: January 1, 1997, 00:00 to:
December 31, 2080, 23:59
Accuracy better than ±1 min. per month
Hold Time infinite if powered by AC; otherwise at least
48 hours