21Paper Save Mode for Maternal
Measurements
Your monitor's recorder features a Paper Save Mode, where maternal vital signs are recorded using less
paper than during a normal trace recording.
When Paper Save Mode is enabled, and if the recorder is stopped, it will start automatically to print
data from maternal measurements as they occur, and then stop again to save paper. You enable Paper
Save Mode in Configuration Mode (default is off).
A header is printed first before the measurements are recorded. A new header is also printed when
there is a date change at midnight.
Each NBP measurement is recorded. The time when the measurement ended is recorded.
Other maternal parameters (SpO
2
, maternal heart rate or Pulse) are recorded every five minutes.
The rules described in the section “Priority for Maternal Heart / Pulse Rate” on page 155 apply.
Paper Save Mode recording stops if there are no valid maternal measurements for more than one
hour, and a message will notify you that there are no active parameters. Paper Save Mode recording
will restart automatically when another valid measurement is made.