26Specifications and Standards
Compliance
The monitors are intended to monitor a mother and her fetus(es), which from an electrical safety point
of view, are one person.
Environmental Specifications
The monitor may not meet the given performance specifications if stored and used outside the
specified temperature and humidity ranges.
Monitor (M2702A/M2703A); Interface Cable for Avalon CTS (M2731-60001 and M2732-60001)
Temperature Range Operating without battery option: 0°C to 45°C (32°F to 113°F)
with battery option/charging: 0°C to 35°C (32°F to
95°F)
with battery option/fully charged: 0°C to 40°C (32°F
to 104°F)
Storage -20°C to 60°C (-4°F to 140°F)
Humidity Range Operating <95% relative humidity @ 40°C/104°F
Storage <90% relative humidity @ 60°C/140°F
Altitude Range Operating -500 to 3000 m/-1640 to 9840 ft.
Storage -500 to 13100 m/-1640 to 43000 ft.
Transducers (M2734A/M2734B/M2735A/M2736A/M2738A)
Temperature Range Operating 0°C to 40°C (32°F to 104°F)
Storage -20°C to 60°C (-4°F to 140°F)
Humidity Range Operating <95% relative humidity @ 40°C/104°F
Storage <90% relative humidity @ 60°C/140°F
Altitude Range Operating -500 to 3000 m/-1640 to 9840 ft.
Storage -500 to 13100 m/-1640 to 43000 ft.