26 Specifications and Standards Compliance
197
Noninvasive Blood Pressure
Complies with IEC 60601-2-30:1999/EN60601-2-30:2000.
Fetal / Maternal Defaults Settings
FHR (Ultrasound/DECG) Alarms On/Off Default On
Default Color for FHR Numeric Orange
Toco Default color for Toco numeric Green
IUP Default IUP Scale Unit mmHg
Default color for IUP numeric Green
Maternal Heart Rate
(MHR) Measurement
Default Color for MECG Numeric Red
Performance Specifications
Measurement
Ranges
Systolic 30 to 270 mmHg (4 to 36 kPa)
Diastolic 10 to 245 mmHg (1.5 to 32 kPa)
Mean 20 to 255 mmHg (2.5 to 34 kPa)
Accuracy Max. Std. Deviation:8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Range 40 to 300 bpm
Accuracy
(average over
noninvasive blood
pressure measurement
cycle)
40 to 100 bpm: ±5 bpm
101 to 200 bpm: ±5% of reading
201 to 300 bpm: ±10% of reading
Measurement Time Typical at HR > 60bpm
Auto/manual: 30 seconds (adult)
Maximum time: 180 seconds (adult)
Cuff Inflation Time Typical for normal adult cuff: Less than 10 seconds
Initial Cuff Inflation Pressure 165 ±15 mmHg
Auto Mode Repetition Times 1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes
Venipuncture Mode Inflation
Inflation Pressure 20 to 120 mmHg (3 to 16 kPa)
Automatic deflation after 170 seconds