Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The Philips Avent breast pump uses RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Philips Avent breast pump is suitable for
use in all establishments, including domestic
establishments and those directly connected to
the public low voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
uctuations/ icker
emissions IEC
61000-3-3
Complies
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specied
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IMMUNITY test IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for
power supply
lines
± 2 kV for
power supply
lines
Mains power quality should be
that of a typical residential or
hospital environment.
SurgeIEC 61000-4-5 ± 1 kV line(s) to
line(s)
± 1 kV line(s)
to line(s)
Mains power quality should be
that of a typical residential or
hospital environment.