Chapter 5 Viewing and Changing Settings
45
Therapy Settings
Mode
Change the Mode setting to one of the following therapy modes:
- CPAP
- S
- S/T
- T
- PC
- AVAPS-AE
Trigger Type
The device can be set to trigger breaths based on automatic ow thresholds or specic ow
settings. Change the Trigger Type to one of the following options: Auto-Trak, Auto-Trak
[Sensitive] or Flow Trigger.
If Flow Trigger is selected then two set points will be available for adjustment: Flow Trigger
Sensitivity and Flow Cycle Sensitivity.
Flow Trigger Sensitivity (Expiration to Inspiration)
The Flow Trigger Sensitivity may be adjusted from 1 to 9 l/min in 1 l/min
increments. The ow trigger initiates when the patient’s inspiratory eort creates a
ow equal to or greater than the ow sensitivity setting.
Flow Cycle Sensitivity (Inspiration to Expiration)
The Flow Cycle Sensitivity may be adjusted from 10 to 90% in 1% increments. As
ow begins to decrease during inspiration, if the patient ow is less than the ow
cycle set point, the device will cycle to expiration.
CPAP
Increase or decrease the CPAP pressure setting from 4 to 20 cm H
2
O in increments of 0.5.
Flex Lock
Select O to allow users to adjust the Flex setting. Or, select On so users cannot adjust their
Flex setting.