Chapter 5 Viewing and Changing Settings
45
Therapy Settings
Mode
Change the Mode setting to one of the following therapy modes:
- CPAP
- S
- S/T
- T
- PC
- AVAPS-AE
Trigger Type
The device can be set to trigger breaths based on automatic ow thresholds or specic ow
settings. Change the Trigger Type to one of the following options: Auto-Trak, Auto-Trak
[Sensitive] or Flow Trigger.
If Flow Trigger is selected then two set points will be available for adjustment: Flow Trigger
Sensitivity and Flow Cycle Sensitivity.
• Flow Trigger Sensitivity (Expiration to Inspiration)
The Flow Trigger Sensitivity may be adjusted from 1 to 9 l/min in 1 l/min
increments. The ow trigger initiates when the patient’s inspiratory eort creates a
ow equal to or greater than the ow sensitivity setting.
• Flow Cycle Sensitivity (Inspiration to Expiration)
The Flow Cycle Sensitivity may be adjusted from 10 to 90% in 1% increments. As
ow begins to decrease during inspiration, if the patient ow is less than the ow
cycle set point, the device will cycle to expiration.
CPAP
Increase or decrease the CPAP pressure setting from 4 to 20 cm H
2
O in increments of 0.5.
Flex Lock
Select O to allow users to adjust the Flex setting. Or, select On so users cannot adjust their
Flex setting.