Changing Alarm Settings
4-12 EarlyVue VS30 Instructions for Use
To automatically set alarm limits
1 To establish a baseline, take an initial set of vital signs measurements on the patient.
2 Touch the Limits button in the toolbar
The Alarm Settings menu opens, displaying the General tab.
3 Touch the Auto Set Alarm Limits button.
A confirmation window appears.
4 Touch OK.
The monitor calculates new high and low alarm limits based on the baseline
measurements.
Note
The alarm limits change for existing measurements only. The alarm limits do not
change if a measurement was not taken.
Restoring Default Alarm Settings
Notes:
• Default alarm settings can be either factory-set defaults or defaults set by your
system administrator. The default values set by your system administrator override
the factory defaults.
• The system automatically restores default values when you admit a new patient,
change the patient type, put the monitor in Standby mode, or shut the monitor down.
The default settings are based on the current patient type.
To restore the default alarm settings
1 Touch the Limits button in the toolbar.
The Alarm Settings menu opens, displaying the General tab.
2 Touch the Restore Default Alarm Settings button.
A confirmation message appears.
3 Touch OK.
The monitor restores all alarm settings to the default values.
For a list of factory default alarm values, see Appendix A, “Alarm Specifications.”