EarlyVue VS30 Instructions for Use 6-11
Manually Starting an NBP Measurement
If an NBP Auto Interval or NBP Interval Program measurement is in progress, manually
taking an NBP may cause the monitor to skip one measurement in the interval sequence,
but the interval sequence will resume after the skipped measurement.
WARNING
Select the correct patient type for your patient. Do not apply the higher adult inflation limits and
measurement duration to neonatal patients. Patient injury may occur if adult inflation pressures are
used for pediatric or neonatal patients. Over-inflating the cuff may cause patient injury, patient
discomfort or inaccurate readings.
Caution
Before starting an NBP measurement, ensure that you have selected the appropriate cuff and
applied it properly. See “Selecting an NBP Cuff” on page 6-2.
NBP menu You can start and stop any type of NBP
measurement from the NBP menu.
Interval and Continuous Monitoring Modes
Interval Settings
menu and
Continuous
Settings menu
If you have set an NBP Auto Interval or
NBP Interval Program, touching Start
Interval starts an NBP measurement.
If you have not set an NBP Auto Interval
or NBP Interval Program, touching Start
STAT starts a STAT NBP measurement.
Location Button Notes
Start QuickNBP
Stop QuickNBP
Start STAT
Stop STAT
Start NBP
Stop NBP
Start Interval
Start STAT