EarlyVue VS30 Instructions for Use 9-3
Connecting the Sampling Line
Note
When using an Infant/Neonatal VitaLine™
H Set with an incubator, position one Nafion
®
humidity filter inside the incubator and the second Nafion humidity filter outside the
incubator, close to the incubator tubing access door.
The maximum CO
2
response time (with a standard-length FilterLine™) is 5.1 seconds
(typical).
Note
Adult/Pediatric Long FilterLines and Long Smart CapnoLines have a maximum CO
2
response time of 6.9 seconds (typical).
For additional information on CO
2
accessories, see Chapter 15, “Approved Accessories
and Sensors.”
Connecting the Sampling Line
Note
The awRR Apnea Alarm timer begins when the sampling line is connected to the monitor
and the CO
2
Hardware option is set to On.
To prepare for connecting the sampling line
1 Ensure that the monitor is in Interval or Continuous monitoring mode.
2 Touch the CO2 numeric pane to open the CO2 menu.
3 Touch the CO2 Hardware menu.
4 Touch On.
5 Touch Done to return to the main screen.