The Resp Numeric Pane
11-2 EarlyVue VS30 Instructions for Use
Summary of Respiration Rate Sources and Entry Methods
The following table summarizes the conditions under which automatic respiration
sources are available, and when you can or must use manual entry methods. Conditions
include the current monitoring mode (left column) and the CO
2
and Masimo RRa options
configured on your monitor (top row),
The Resp Numeric Pane
The following illustration shows the components of the Resp numeric pane:
When your monitoring mode and optional modules do not support an automatically
calculated respiration rate, you can manually enter the respiration rate directly in the
Resp numeric pane. See “To enter a manual respiration rate from the Resp pane” on
page 11-6.
No CO
2
and no
RRa
CO
2
only (no RRa) RRa only (no CO
2
)CO
2
and RRa
Spot Check
mode
Manual entry on
main screen only
Manual entry on
main screen only
Automatic RRa or
manual entry in
Respiration menu
Automatic RRa or
manual entry in
Respiration menu
Interval or
Continuous
mode
Manual entry on
main screen only
Automatic awRR or
manual entry in
Respiration menu
Automatic RRa or
manual entry in
Respiration menu
Automatic RRa or
awRR or manual
entry in
Respiration menu
20
High and low alarm limits
Respiration rate source (awRR in this
example)
Respiration rate value
rpm
Source awRR
Unit of measure
30
8
Resp