Using Interval and Continuous Monitoring Mode with EWS and IGS
12-10 EarlyVue VS30 Instructions for Use
Notes:
You can view the EWS recommended actions again by opening the record and
touching the EWS button at the lower right of the expanded record.
If the export to the EMR system fails, the indicator turns blue. See “Patient Record
Colors” on page 3-16
Using Interval and Continuous Monitoring Mode with EWS and IGS
If you are operating in Interval or Continuous monitoring mode (unattended records), you
can still use EWS with IGS to take an attended measurement and display an Early
Warning Score (EWS). The score is sent to IGS.
During Interval or Continuous monitoring, early warning scores are not calculated when
vital signs measurements are taken automatically. To generate early warning scores, you
must touch the Save Now button on the main screen while you are at the bedside.
For general information about Interval and Continuous monitoring modes, see
“Monitoring Mode Menu” on page 2-18.
Troubleshooting IGS
Use the following table to diagnose and fix IGS problems. The table describes the
problem by symptom, lists possible causes and suggests an action. If you cannot resolve
a problem by using the following table, or if the table does not list the problem you are
experiencing, see your system administrator.
Symptom Possible Cause Action
No EWS score
displayed
Incomplete measurement data. Find out from your system
administrator which vitals are
required.
Your system administrator can
configure which vital signs are
required for EWS analysis.
Take all vitals and provide all
required observations and
assessments before saving
the patient record.
The date/time on the monitor
is not current and is not
synchronized with the IGS
server.
Ask your system
administrator whether the
date and time settings are
configured properly.