EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
79
Access to E-Module
The E-Module contains most of the active circuitry of the system.
To open the E-Module:
Locate the peripheral bay release mechanisms on each side of
system.
Slide release mechanism towards back of system.
Lift the peripheral bay up and back.
Remove card cage cover using two rotating locks.