2-4
PHILIPS MEDICAL SYSTEMS
NOTE: Be careful not to overpack the carrying case, to avoid placing
inadvertent pressure on the control buttons. Do not store the FR2+ with
the defibrillator pads attached. Do not open the pads package until ready
for use.
With the battery installed and the FR2+ stored in appropriate environmental
conditions,
*
the FR2+ performs detailed daily, weekly, and monthly selftests to
check its readiness for use. These periodic selftests are described in Chapter 4.
While the FR2+ is in the standby mode, the Status Indicator shows the flashing
black hourglass unless the periodic selftest detects a problem. If a problem is
detected, the Status Indicator will show a flashing red X or a solid red X and the
FR2+ will chirp to alert you. Chapter 4 contains instructions for
troubleshooting.
* See Appendix B for environmental specifications.