AFTER USING THE HEARTSTART FR2+
After each use of the FR2+, in addition to the maintenance tasks described in
the table above, perform the following post-use checks before returning the
FR2+ to service:
• Check the operation of the FR2+ by removing and reinstalling the battery and
running the battery insertion selftest.
• Check the outside of the FR2+ and the connector socket for signs of dirt or
contamination. If the FR2+ is dirty or contaminated, clean it according to the
guidelines provided in this manual.
daily monthly maintenance task/response
Check the Status Indicator.
If you see the flashing black hourglass: The FR2+ is
ready to use. No action required.
If you see anything other than a flashing black hourglass,
remove and reinstall the battery to run the selftest.
• If the selftest passes and the Status Indicator shows
the flashing black hourglass, the FR2+ is ready to use.
• If the selftest fails, install a new battery and run the
selftest. If the selftest passes, the FR2+ is ready to
use. If the selftest fails, contact Philips Medical
Systems.
• If the selftest does not run, check to be sure that
there is no pads connector installed in the FR2+.
Check supplies, accessories, and spares for damage and
expiration dating.
Do not use damaged or expired accessories. Replace
them immediately.
If a LOW BATTERY or REPLACE BATTERY message is
displayed: Replace the battery and run the selftest.
• Do not attempt to charge the M3863A FR2 standard
battery. It is not rechargeable.
• The M3848A FR2+ battery is rechargeable. Recharge
it, using the M3849A Charger only.
Check the outside of the FR2+ and the connector
socket for cracks or other signs of damage.
If you see signs of damage: Contact Philips for technical
support.