4-2
PHILIPS MEDICAL SYSTEMS
AFTER USING THE HEARTSTART FR2+
After each use of the FR2+, in addition to the maintenance tasks described in
the table above, perform the following post-use checks before returning the
FR2+ to service:
Check the operation of the FR2+ by removing and reinstalling the battery and
running the battery insertion selftest.
Check the outside of the FR2+ and the connector socket for signs of dirt or
contamination. If the FR2+ is dirty or contaminated, clean it according to the
guidelines provided in this manual.
daily monthly maintenance task/response
Check the Status Indicator.
If you see the flashing black hourglass: The FR2+ is
ready to use. No action required.
If you see anything other than a flashing black hourglass,
remove and reinstall the battery to run the selftest.
If the selftest passes and the Status Indicator shows
the flashing black hourglass, the FR2+ is ready to use.
If the selftest fails, install a new battery and run the
selftest. If the selftest passes, the FR2+ is ready to
use. If the selftest fails, contact Philips Medical
Systems.
If the selftest does not run, check to be sure that
there is no pads connector installed in the FR2+.
Check supplies, accessories, and spares for damage and
expiration dating.
Do not use damaged or expired accessories. Replace
them immediately.
If a LOW BATTERY or REPLACE BATTERY message is
displayed: Replace the battery and run the selftest.
Do not attempt to charge the M3863A FR2 standard
battery. It is not rechargeable.
The M3848A FR2+ battery is rechargeable. Recharge
it, using the M3849A Charger only.
Check the outside of the FR2+ and the connector
socket for cracks or other signs of damage.
If you see signs of damage: Contact Philips for technical
support.