• Check the data card if one has been used. If the data card has been used to
record incident data, remove and replace it with a blank data card. Deliver the
recorded data card to appropriate personnel according to your local
guidelines and medical protocol.
• Check the connector socket to make sure that defibrillator pads are
disconnected from the FR2+ when it is not in use.
• Check to make sure the data card tray is installed, even if a data card is not
being used.
CLEANING THE HEARTSTART FR2+
The outside of the FR2+, including the defibrillator pads connector socket, can
be cleaned with a soft cloth dampened in one of several appropriate cleaning
agents (see list below). The following guidelines include some important
reminders:
• Do not immerse the FR2+ in fluids.
• Make sure a battery (or the M3864A Training & Administration Pack) and a
data card tray are installed when cleaning the FR2+, to keep fluids out of the
device.
• Do not use abrasive materials, cleaners, strong solvents such as acetone or
acetone-based cleaners, or enzymatic cleaners.
• Clean the FR2+ and the connector socket with a soft cloth dampened with
one of the cleaning agents listed below.
— Isopropyl alcohol (70% solution)
—Soapy water
— Chlorine bleach (30 ml/l water)
—Ammonia-based cleaners
— Glutaraldehyde-based cleaners
— Hydrogen peroxide
CAUTION:Do not immerse any portion of the FR2+ in water or other
fluids. Do not allow fluids to enter the FR2+. Avoid spilling any fluids on the
FR2+ or accessories. Spilling fluids into the FR2+ may damage it or present a
fire or shock hazard. Do not sterilize the FR2+ or accessories.