4-3
PHILIPS MEDICAL SYSTEMS
4
Check the data card if one has been used. If the data card has been used to
record incident data, remove and replace it with a blank data card. Deliver the
recorded data card to appropriate personnel according to your local
guidelines and medical protocol.
Check the connector socket to make sure that defibrillator pads are
disconnected from the FR2+ when it is not in use.
Check to make sure the data card tray is installed, even if a data card is not
being used.
CLEANING THE HEARTSTART FR2+
The outside of the FR2+, including the defibrillator pads connector socket, can
be cleaned with a soft cloth dampened in one of several appropriate cleaning
agents (see list below). The following guidelines include some important
reminders:
Do not immerse the FR2+ in fluids.
Make sure a battery (or the M3864A Training & Administration Pack) and a
data card tray are installed when cleaning the FR2+, to keep fluids out of the
device.
Do not use abrasive materials, cleaners, strong solvents such as acetone or
acetone-based cleaners, or enzymatic cleaners.
Clean the FR2+ and the connector socket with a soft cloth dampened with
one of the cleaning agents listed below.
Isopropyl alcohol (70% solution)
—Soapy water
Chlorine bleach (30 ml/l water)
—Ammonia-based cleaners
Glutaraldehyde-based cleaners
Hydrogen peroxide
CAUTION:Do not immerse any portion of the FR2+ in water or other
fluids. Do not allow fluids to enter the FR2+. Avoid spilling any fluids on the
FR2+ or accessories. Spilling fluids into the FR2+ may damage it or present a
fire or shock hazard. Do not sterilize the FR2+ or accessories.