4-9
PHILIPS MEDICAL SYSTEMS
4
BATTERY READINESS — a GOOD BATTERY message (M3863A) or a fuel
gauge display (M3848A) showing 25%, 50%, 75% or 100%, or a
LOW BATTERY
or
REPLACE BATTERY message, as appropriate.
STATUS — the current status of the M3863A battery. Make a note of this
code if technical service is needed.
TROUBLESHOOTING GUIDE
STATUS INDICATOR SUMMARY
RECOMMENDED ACTION DURING AN EMERGENCY
If the status indicator displays the flashing black hourglass, follow all voice and
text prompts.
The HeartStart FR2+ Defibrillator is designed to deliver therapy even if the
status indicator displays a flashing red X, although the device may not perform
to all of its specifications. Voice and text prompts should be followed whenever
they are given. If for any reason you cannot hear voice prompts during use of the
defibrillator, periodically check the device screen for text prompts.
NOTE:After completing emergency use of the FR2+, if you are unable to
clear the problem as described in this Troubleshooting section, and the
Status Indicator does not show the flashing black hourglass, contact Philips
for technical support.
status indicator meaning
Flashing black hourglass The FR2+ passed the battery insertion selftest
or the last periodic selftest and is therefore
ready for use.
Flashing red X
accompanied by a chirping
sound.
A selftest error has occurred or the battery
power is low.
Solid red X The battery is completely depleted or a selftest
failure occurred.