4-10
PHILIPS MEDICAL SYSTEMS
In the unlikely event that the device becomes unresponsive during use:
1. cycle power (press the On/Off button once, wait one second, then press it
again), or
2. remove and reinstall the battery (use a new FR2 standard battery, if
available, or a charged FR2+ rechargeable battery).
If neither of these actions clears the problem, do not use the FR2+. Attend to
the patient, providing CPR if needed, until emergency medical personnel arrive.
NOTE:Perform CPR (if needed) any time there is a delay before or an
interruption in use of the FR2+.
TROUBLESHOOTING DURING PATIENT USE
symptom possible cause recommended action
STATUS INDICATOR: FLASHING RED X
Screen and voice prompts:
LOW BATTERY
Screen and voice prompts:
REPLACE BATTERY NOW
• The energy remaining in the
battery is low but sufficient to
deliver a shock.
• The energy in the battery is nearly
depleted. The FR2+ will turn off if a
new battery is not installed.
Replace the battery with a new FR2
standard or a charged FR2+
rechargeable battery as soon as
possible.
STATUS INDICATOR: FLASHING BLACK HOURGLASS
Screen and voice prompts:
APPLY PADS
PRESS PADS FIRMLY
or
PLUG IN CONNECTOR
Or voice prompts:
INSERT CONNECTOR
FIRMLY
or
PRESS PADS FIRMLY TO
PATIENT'S BARE CHEST
or
POOR PADS CONTACT
• The defibrillator pads are not
properly applied to the patient, or
• The pads are not making good
contact with the patient's bare
chest because of moisture or
excessive hair, or
• The pads are touching each other,
or
• The defibrillator pads connector is
not firmly inserted in the
connector socket.
• Make sure that the defibrillator
pads are sticking completely to
the patient’s skin.
• If the pads are not sticking, dry
the patient's chest and shave or
clip any excessive chest hair.
• Reposition the pads.
• Make sure the pads connector is
completely inserted in the
connector socket.
If the prompt continues after you
do these things, replace the pads.