Voice and text prompts:
REPLACE PADS
• The defibrillator pads, cable, or
connector may be damaged.
• The FR2+ has detected a possible
problem with the defibrillator pads
or pads cable.
Replace the defibrillator pads with
new defibrillator pads.
Voice promp ts:
ANALYZING
INTERRUPTED
or
STOP ALL MOTION
or
CANNOT ANALYZE
• The patient is being moved or
jostled.
• Radio or electrical sources are
interfering with ECG analysis.
• The environment is dry and
movement around the patient is
causing static electricity to
interfere with ECG analysis.
• Stop CPR; do not touch the
patient. Minimize patient motion.
If the patient is being
transported, stop the vehicle if
needed.
• Check for possible causes of
radio and electrical interference
and remove them from the area.
• Responders and bystanders
should minimize motion,
particularly in dry environments
that can generate static
electricity.
Voice and text prompts:
NO SHOCK DELIVERED
The patient impedance is not within
the specifications for the FR2+ to
deliver a shock.
• Make sure the defibrillator pads
are correctly positioned on the
patient according to the diagram
on the back of the pads.
• Make sure the defibrillator pads
connector is completely inserted
in the connector socket.
• Press the defibrillator pads firmly
to the patient's chest.
• Replace the defibrillator pads if
necessary.
Voice prompt:
SHOCK BUTTON NOT
PRESSED
Shock has been advised but not
delivered within 30 seconds. (FR2+
has been disarmed.)
When next prompted, press the
Shock button to deliver shock.
symptom possible cause recommended action