4-11
PHILIPS MEDICAL SYSTEMS
4
Voice and text prompts:
REPLACE PADS
The defibrillator pads, cable, or
connector may be damaged.
The FR2+ has detected a possible
problem with the defibrillator pads
or pads cable.
Replace the defibrillator pads with
new defibrillator pads.
Voice promp ts:
ANALYZING
INTERRUPTED
or
STOP ALL MOTION
or
CANNOT ANALYZE
The patient is being moved or
jostled.
Radio or electrical sources are
interfering with ECG analysis.
The environment is dry and
movement around the patient is
causing static electricity to
interfere with ECG analysis.
Stop CPR; do not touch the
patient. Minimize patient motion.
If the patient is being
transported, stop the vehicle if
needed.
Check for possible causes of
radio and electrical interference
and remove them from the area.
Responders and bystanders
should minimize motion,
particularly in dry environments
that can generate static
electricity.
Voice and text prompts:
NO SHOCK DELIVERED
The patient impedance is not within
the specifications for the FR2+ to
deliver a shock.
Make sure the defibrillator pads
are correctly positioned on the
patient according to the diagram
on the back of the pads.
Make sure the defibrillator pads
connector is completely inserted
in the connector socket.
Press the defibrillator pads firmly
to the patient's chest.
Replace the defibrillator pads if
necessary.
Voice prompt:
SHOCK BUTTON NOT
PRESSED
Shock has been advised but not
delivered within 30 seconds. (FR2+
has been disarmed.)
When next prompted, press the
Shock button to deliver shock.
symptom possible cause recommended action