4-13
PHILIPS MEDICAL SYSTEMS
4
Status Indicator:
FLASHING OR SOLID
RED X
Audio signal:
CHIRPING
Text prompt (displayed for
10 seconds at the end of a
BIT, before FR2+ turns off):
NOT READY FOR USE
or
SELFTEST FAILED
A test revealed a failure or error. The
FR2+ performs selftests every time it
is turned on, when a battery is
inserted, and periodically while it is in
standby mode.
Unplug the pads connector from
the FR2+, if connected.
Remove and reinstall the battery
and check the results of the
battery insertion selftest. If it
fails, install a new FR2 standard
battery or a charged FR2+
rechargeable battery and repeat
the test. If it fails again, do not
use the FR2+.
NOTE: You can stop the tests and use
the FR2+ as soon as you see the Status
Indicator change to the flashing black
hourglass. Simply press the On/Off
button to stop the test and put the
FR2+ into standby mode. The FR2+ is
then ready for use.
Status Indicator:
SOLID RED X
Audio signal:
NONE
The battery is missing or
completely depleted.
The Training & Administration Pack
is being used in the administration
function (the solid red X is normal
in this case) or has been left in the
FR2+ by mistake.
A selftest detected a failure.
•Install a new FR2 standard battery
or a charged FR2+ rechargeable
battery in the FR2+ and perform
the battery insertion test (BIT).
Remove the Training &
Administration Pack and install a
battery.
Remove and reinstall the battery
and perform the battery
insertion selftest. If it fails, install
a new FR2 standard battery or a
charged FR2+ rechargeable
battery and repeat the test. If it
fails again, do not use the FR2+.
symptom possible cause recommended action