PHILIPS MEDICAL SYSTEMS
Status Indicator:
SOLID RED X
Audio signal:
CHIRPING
• The Training & Administration Pack
is being used in the
ADMINISTRATION function and
more than 10 minutes have passed
without user interaction (button
press or pads change).
• The Training & Administration Pack
is being used in the
TRAINING
function and more than 30 minutes
have passed without user
interaction (button press or pads
change).
• To continue using the Training &
Administration Pack, press any
button (except On/Off).
•To return the FR2+ to standby
mode, remove the Pack and
install a battery.
Status Indicator:
NONE
The FR2+ has been physically
damaged.
• Check for visible damage. Do not
use the FR2+ if it appears to be
damaged.
• Remove and reinstall the battery
to perform the battery insertion
selftest. If it fails, install a new FR2
standard battery or a charged
FR2+ rechargeable battery and
repeat the test. If it fails again, do
not use the FR2+.
symptom possible cause recommended action