5 SAFETY CONSIDERATIONS
You should be aware of the safety concerns listed here when you use the
HeartStart FR2+. Read them carefully. You will also see some of these messages
in other parts of this manual. The messages are labeled DANGER, WARNING,
or CAUTION.
DANGER— immediate hazards that will result in personal injury or
death.
WARNING— conditions, hazards, or unsafe practices that can result
in serious personal injury or death.
CAUTION — conditions, hazards, or unsafe practices that can result in
minor personal injury, damage to the HeartStart FR2+, or loss of data
stored in the device.
These safety considerations are divided into four groups: safety concerns about
the HeartStart FR2+ in general use, defibrillation, monitoring, and maintenance
activities. Unless otherwise noted, the dangers, warnings, and cautions listed in
the following tables apply to the FR2+ M3860A and M3861A.
GENERAL DANGERS, WARNINGS, AND CAUTIONS
safety level possible shock or fire hazard, or explosion
DANGER THERE IS A POSSIBILITY OF EXPLOSION IF THE HEARTSTART FR2+ IS USED IN
THE PRESENCE OF FLAMMABLE ANESTHETICS OR CONCENTRATED
OXYGEN. MOVE SUPPLEMENTAL OXYGEN AND OXYGEN DELIVERY DEVICES
AWAY FROM THE DEFIBRILLATION PADS. HOWEVER, IT IS SAFE TO USE THE
FR2+ ON SOMEONE WEARING AN OXYGEN MASK.
DANGER THE HEARTSTART FR2+ HAS NOT BEEN EVALUATED OR APPROVED FOR USE
IN HAZARDOUS LOCATIONS AS DEFINED IN THE NATIONAL ELECTRICAL
CODE (ARTICLES 500-503). IN ACCORDANCE WITH THE IEC
CLASSIFICATIONS (SECTION 5.5.), THE HEARTSTART FR2+ IS NOT TO BE
USED IN THE PRESENCE OF FLAMMABLE SUBSTANCE/AIR MIXTURES.
DANGER DO NOT RECHARGE THE M3863A FR2 STANDARD BATTERY.
WARNING Use the HeartStart FR2+ only as described in this manual. Improper use of the
HeartStart FR2+ can cause death or injury. Do not press the Shock button if the
defibrillator pads are touching each other or are open and exposed.