MANUAL OVERRIDE PARAMETERS
The parameters in the following table are used to enable different kinds of
manual override.
parameter settings default description
advanced OFF,
ANALYZE,
CHARGE
OFF Enables or disables advanced mode entry for ALS or tiered-response
systems.
OFF — disables advanced mode features.
ANALYZE — enables user-initiated rhythm analysis and disarm, and
(M3860A only) automatically turns on ECG display when advanced
mode is entered.
CHARGE (M3860A only) — in addition to enabling the analyze feature,
enables user-initiated charging and disarming.
pause key ON, OFF OFF Enables (ON) or disables (OFF) a user-initiated CPR interval in the
automatic protocol. The interval length is defined by the CPR timer
setting. The pause key is disabled when an advanced mode feature
(analyze or charge) is enabled and accessed, and during Monitoring.
OFF — disables availability of user-initiated pause.
ON — enables user-initiated pause by pressing the lower Option button
indicated by an arrow on the FR2+ display, at any time except when the
device is monitoring or is already paused.
If enabled, the pause is initiated by
pressing the lower Option button
indicated by an arrow on the FR2+
display, as shown in the sample screen.