6-6
PHILIPS MEDICAL SYSTEMS
USING SETUP FEATURES
NOTE:To move around the menus displayed, use the Option buttons as
follows:
•Press the lower Option button to move the highlight bar from one item
to another on the menu.
•Press the upper Option button to select the highlighted item or to scroll
through the settings for that item.
The FR2+ comes with a factory default setup designed to meet the needs of
most users. The setup feature of the FR2+ lets you review the current setup of
your HeartStart FR2+ or install a revised setup if appropriate. To go to the
SETUP menu:
1. Remove and reinstall the battery to bring up the first menu on the screen.
resume key ON, OFF OFF Enables (ON) or disables (OFF) user-initiated interruption of CPR and
patient care intervals and return to analyzing, by pressing the lower
Option button indicated by an arrow on the FR2+ display. If either the
CPR timer or the NSA action setting is programmed to 2.5 minutes or
longer, the Resume Key setting is automatically enabled (ON). The
Resume Key is always automatically enabled for any CPR First interval.
If enabled, analysis is initiated by pressing the lower Option button
indicated by an arrow on the FR2+ display, as shown in the sample screen:
advanced use
prompt
interval
(minutes)
0.5, 1.0,
1.5, 2.0,
2.5, 3.0
0.5 Sets the interval for patient care prompts provided during advanced mode
operation.
parameter settings default description