6-7
PHILIPS MEDICAL SYSTEMS
6
NOTE:This screen will not be displayed if the FR2+ is connected to
defibrillator pads (that are applied to the patient) when the battery is
inserted, and you will not be able to access the menu items. In addition, the
battery insertion selftest and periodic automatic selftests cannot run while
the defibrillator pads are connected. Be sure to unplug the pads connector from
the FR2+ after each use. Do not store the FR2+ with the pads connected.
2. Within 10 seconds of installing the battery, press the lower Option button
to move the highlight bar to NEXT.
3. Press the upper Option button to select NEXT.
4. Press the lower Option button to move the highlight bar to SETUP.
5. Press the upper Option button to bring up the SETUP menu.
The SETUP menu allows you to receive setup directly from another HeartStart
FR2+, read setup from a data card, or review current setup.
REVIEWING CURRENT SETUP
A good way to understand the setup of your FR2+ is to review the setup it
currently uses.
1. Select REVIEW SETUP from the SETUP menu. The first of a series of
REVIEW SETUP screens is displayed.
2. After reviewing the screen contents, press the upper Option button to
select NEXT and move to the next screen.
3. The last screen allows you to select RETURN and go back to the SETUP
menu.