6-13
PHILIPS MEDICAL SYSTEMS
6
If the rhythm analysis results in a Shock Advised decision, the FR2+ begins
charging, prompts you to press the Shock button, and displays a MANUAL
DISARM option at the top of the screen. If for any reason you want to cancel
the shock, press the upper Option button to disarm the FR2+.
To return to non-manual, AED mode operation, turn the FR2+ off by pressing
the On/Off button. Then turn the FR2+ on by pressing the On/Off button again.
USING THE MANUAL CHARGE FEATURE (M3860A ONLY)
The manual charge feature is available only in the M3860A, when enabled in
setup.
To enter the advanced mode during use of an FR2+ that has this feature enabled,
make sure pads are attached to the patient, then press both Option buttons
simultaneously. This brings up a screen that includes a highlighted line at the top,
labeled MANUAL, with an arrowhead pointing to the upper Option button, and
another at the bottom, labeled ANALYZE, with an arrowhead pointing to the
lower Option button.
When the advanced mode is entered, display of the patient’s ECG and heart rate
is automatically initiated.
Pressing the lower Option button (ANALYZE) provides user-initiated rhythm
analysis as described above. Pressing the upper Option button (MANUAL)
brings up a new screen.
The highlighted top line is labeled MANUAL CHARGE, with an arrowhead
pointing to the upper Option button.
If the ECG display shows that, in your expert clinical judgment, the patient has a
shockable rhythm, press the upper Option button (MANUAL CHARGE). The
HeartStart FR2+ will immediately charge for shock delivery.
As soon as charging begins, the screen message changes to CHARGING,
STAND CLEAR, and the label for the arrowhead pointing to the upper Option
button changes to MANUAL DISARM.