7-4
PHILIPS MEDICAL SYSTEMS
5. To review the events that occurred during the incident, select REVIEW
EVENTS. A new screen comes up. This and following screens, accessed by
selecting NEXT EVENTS, display elapsed time information for critical
activities in using the FR2+. These include:
• POWER ON — when the FR2+ was turned on,
• PADS ON — when the defibrillator pads were connected,
• SHOCK ADVISED — when a shock was advised,
• ARMED — when the FR2+ is fully charged for shock delivery,
• SHOCKED — when a shock was delivered,
• PAUSE FOR CPR — when a pause occurred
• POWER OFF — when the FR2+ was turned off
Additional information may be displayed if your FR2+ is using a revised
setup allowing advanced mode operation.
6. To review the first six seconds of the recorded ECG for the incident, select
REVIEW ECG. A new screen comes up. This screen displays a three-second
segment of the presenting ECG from the incident.
7. Select NEXT ECG SEGMENT to review the second three-second segment
of the presenting ECG.
Data cards can be reused if desired. Using a personal computer running
HeartStart Event Review software and an appropriate card reader for your
computer, you can copy the information from a data card, then erase the card
and reuse it in the FR2+.