7-3
PHILIPS MEDICAL SYSTEMS
7
4. Observe and record, if desired, the summary information displayed on the
screen:
• how long the incident recorded by the FR2+ lasted, and
• how many shocks were delivered during the incident.
This information remains in the FR2+’s memory and can be displayed for review
until the next time the FR2+ is used. At that time, the data from the new
incident will be displayed. Summary data for the most recent use is always saved
in the FR2+ memory until the FR2+ is used again.
REVIEWING DATA FROM A DATA CARD
If a data card is installed when the HeartStart FR2+ is turned on for use during
an incident, the HeartStart FR2+ automatically records detailed information on
the data card. To review this information on the HeartStart FR2+ screen:
1. Make sure the Training & Administration Pack is not installed.
2. Make sure the data card is installed. Unplug the pads connector.
3. Remove and reinstall the battery.
4. Select REVIEW INCIDENT from the menu. A new screen comes up. This
screen displays:
• ELAPSED TIME — how long the incident recorded by the FR2+ lasted,
*
• SHOCKS DELIVERED — how many shocks were delivered during the
incident, and
• FIRST SHOCKS AT — the times at which the first three shocks were
delivered.
NOTE: If the data card does not contain event data, only the summary
information from FR2+ internal memory will be displayed when REVIEW
INCIDENT is selected.
* The FR2+ displays elapsed time to a maximum of 99:59 minutes. If the elapsed time of
use extends beyond this figure, the minutes are represented by “??” but the seconds
are displayed. However, total elapsed time will be recorded on an installed data card
for later review with HeartStart Event Review data management software.