7-2
PHILIPS MEDICAL SYSTEMS
and remove the card. Give it to the appropriate person in your organization.
Either load a new data card into the tray or insert the empty data card tray into
the port.
NOTE: Because it helps seal the FR2+ against moisture, the data card tray
should always be reinserted into the port of the FR2+.
To avoid running out of data card space during an incident, it is recommended
that each data card be used to record the information for only one incident and
that it be replaced, or erased and reinstalled, after each use of the FR2+.
If you record information from more than one incident on a data card, it is
important to review how much time is left on the used data card before
recording a new incident. To do this, load the data card into the data card tray,
insert the tray in the FR2+, then remove and reinstall the battery. The first
screen displayed shows how much recording time remains on the card.
NOTE:During an incident, if for any reason you turn off the FR2+ for less
than five minutes, the FR2+ considers this to be a “continued use” situation,
and:
the information stored about the incident is saved,
additional events recorded after the device is turned back on will be
treated as part of the same incident, and
the selftest will not automatically run if the battery is replaced.
IMPORTANT: Do not remove the battery while incident data are being
recorded to a data card. To ensure that no incident data are lost, turn the
FR2+ off (return it to standby mode) before replacing the battery.
REVIEWING INCIDENT DATA
REVIEWING DATA FROM INTERNAL MEMORY
Summary information from the last incident that is stored in the internal
memory of the HeartStart FR2+ can be displayed on its screen for review. To
review this information:
1. Remove the data card if one is installed and unplug the pads connector.
2. Remove and reinstall the battery. (Make sure you are using the gray FR2
standard battery or the blue rechargeable battery, not the yellow Training &
Administration Pack.)
3. Select REVIEW INCIDENT from the menu. A new screen comes up.