PHILIPS MEDICAL SYSTEMS
B-8
(OPTIONAL) M3848A FR2+ RECHARGEABLE BATTERY
(OPTIONAL) M3849A CHARGER
maintenance and calibration
requirements for continued
airworthiness
(989803136291 only)
There are no periodic maintenance or calibration requirements that are
necessary for continued airworthiness of the 989803136291 battery. For
battery maintenance with respect to performance within the AED, please see
Chapter 4. There are no user-serviceable parts in the battery.
environmental qualification Meets the applicable requirements of RTCA/DO-227, Section 2.3.
category nominal specifications
category nominal specifications
battery type 12 VDC, 2.2 Ah, lithium ion. Rechargeable cell using the M3849A charger.
capacity When freshly charged and used at 77° F (25° C), provides a minimum of 80
shocks (typically 100 shocks), or 3.5 hours (typically 5 hours) of ECG display
time only, before recharging is indicated.
standby life
(after installation)
6 months when installed fully charged in a defibrillator labeled FR2+.
status indicators Good battery: bar graph on display screen indicating remaining power level.
Low battery: flashing red X on the front panel of the FR2+ (When low
battery indicator appears, there is still enough energy to deliver 9 shocks
plus 15 minutes of ECG display time).
Dead battery: solid red X on the front panel of the FR2+.
storage/transport
temperature
32° to 109° F (0° to 43° C).
category nominal specifications
application For use with M3848A FR2+ rechargeable battery only.
power requirements 100 to 240 VAC, 47 to 63 Hz, 30 Watts
storage/transport
temperature
32° to 122° F (0° to 50° C).
conformance testing International: EN60335-1:1994 Class 1.
North America: UL 1310 Class 2.