PHILIPS MEDICAL SYSTEMS
B
B-9
M3870A AND 989803158211/989803158221 DEFIBRILLATOR PADS
(OPTIONAL) M3854A DATA CARD
(OPTIONAL) M3864A TRAINING & ADMINISTRATION PACK
category nominal specifications
pads, cable, and connector Disposable and self-adhesive. 989803158211/989803158221 adult
defibrillator pads and M3870A FR2 infant/child reduced-energy defibrillator
pads have a minimum active surface area of 85 cm
2
each and are provided in a
sealed package with an integrated 122 cm (48 inch), typical, cable and
connector. The M3870A connector incorporates attenuating electronics.
storage/transport
temperature
32° to 110° F (0° to 43° C).
defibrillator pad
requirements
Use only 989803158211/989803158221, M3870A, M3713A, and M3716A
defibrillator pads with the HeartStart FR2+. Place the pads on the patient as
illustrated on each pad.
category nominal specifications
capacity 8 hours of event and ECG data, or 60 minutes with voice recording.
category nominal specifications
battery type 12 V, 1.1 Ah, nickel metal hydride. Disposable, rechargeable cell using the
M3855A charger.
capacity Provides 4 hours of operating time at 77 °F (25 °C).
status indicators Low battery: flashing red X on the front panel of the FR2+.
Dead battery: solid red X on the front panel of the FR2+.
storage/transport
temperature
50° to 104° F (10° to 40° C).