E-7
PHILIPS MEDICAL SYSTEMS
E
ACCESSORIES SPECIFICATIONS
BATTERY M5070A
HEARTSTART ADULT SMART PADS M5071A AND
INFANT/CHILD SMART PADS M5072A
category specifications
battery type 9 VDC, 4.2 Ah, lithium manganese dioxide. Disposable, long-life primary cell.
capacity When new, a minimum of 200 shocks or 4 hours of operating time at 25° C
(77° F).
shelf life
(prior to insertion)
A minimum of 5 years from date of manufacture when stored and maintained
according to directions provided in this Owner’s Manual.
standby life
(after insertion)
Typically, 4 years when stored and maintained according to directions
provided in this Owner’s Manual.
training life Supports 10 hours of use in training mode.
category specifications
adult pads Disposable, adhesive defibrillation pads with a nominal active surface area of
85 cm
2
each, provided in a snap-in cartridge with an integrated 137.1 cm
(54”), typical, cable.
infant/child pads Disposable, adhesive defibrillation pads with a nominal active surface area of
85 cm
2
each, provided in a snap-in cartridge with an integrated 101.6 cm
(40”), typical, cable. Cartridge incorporates teddy bear icon on cover of seal
for ready identification.
defibrillation pad
requirements
Use only HeartStart Adult SMART Pads M5071A or Infant/Child SMART
Pads M5072A with the HeartStart OnSite Defibrillator.